products

Bante900P Portable pH/Conductivity/Dissolved Oxygen Meter

short description:

Multiparameter portable water quality meter contains the 8 measurement modes, the meter is suitable for measuring the pH, ORP, ion concentration, conductivity and dissolved oxygen.


Product Detail

Product Tags

pH Mode

• Multiparameter water quality meter is equipped with a backlit LCD display.

• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.

• Automatic electrode diagnosis shows the pH slope and offset.

ORP Mode

• 1 point offset calibration allows adjusting the displayed value to a known standard.

• Relative and absolute millivolt modes ensure the reliable ORP measurements.

1

Ion Concentration Mode

• 2 to 5 points calibration, including the 8 concentration points can be selected.

• Direct ion concentration readout simplifies the measurement process.

• Automatic electrode diagnosis shows the calibration points and slopes.

• Selectable concentration units (ppm, mg/L, mol/L, mmol/L) and ionic valency.

Conductivity/TDS/Salinity/Resistivity Modes

• 1 to 5 points calibration with automatic recognition for conductivity standards.

• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and pure water compensations, seawater and practical salinity measurement modes.

Dissolved Oxygen Mode

• 1 or 2 points calibration using the air-saturated water or zero oxygen solution.

• Salinity and barometric pressure compensations eliminate the measurement error.

5

Other Features

• Automatic temperature compensation ensures accurate readings over the entire range.

• Auto-Read function senses and locks the measurement endpoint.

• Calibration due alarm prompts user to calibrate the meter regularly.

• Auto-Power Off effectively conserves battery life.

• Setup menu allows setting the number of calibration points, stability criteria, temperature unit, date and time, etc.

• Reset function automatically resumes all settings back to the factory defaults.

• Expanded memory stores or recalls up to 500 data sets.

• USB communication interface is easy to transfer data to PC.

• Multi-mode power scheme (batteries, power adapter and USB port) ensures that using the meter smoothly.

Specifications

   Model Bante900P
   pH  
   Range -2.000 to 20.000pH
   Resolution 0.001, 0.01, 0.1pH, selectable
   Accuracy ±0.002pH
   Calibration Points 1 to 5 points
   pH Buffer Options USA, NIST, DIN, 2 custom buffers
   ORP  
   Range ±1999.9mV
   Resolution 0.1, 1mV, selectable
   Accuracy ±0.2mV
   Calibration Points 1 point
   Measurement Modes Relative or absolute mV
   Ion Concentration  
   Range 0.001 to 19999 (deponding on the range of ISE)
   Resolution 0.001, 0.01, 0.1, 1
   Accuracy ±0.5% F.S. (monovalent), ±1% F.S. (divalent)
   Calibration Points 2 to 5 points
   Calibration Solutions 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
   Measurement Units ppm, mg/L, mol/L, mmol/L
   Conductivity  
   Range 0.01 to 20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
   Resolution 0.001, 0.01, 0.1, 1
   Accuracy ±0.5% F.S.
   Calibration Points 1 to 5 points
   Calibration Solutions 10µS, 84µS, 1413µS, 12.88mS, 111.8mS
   Temperature Coefficient Linear (0.0 to 10.0%/°C), pure water
   Reference Temperature 20/25°C
   Cell Constant   K=0.1, 1, 10 or custom
   TDS  
   Range 0.00 to 10.00, 100.0, 1000ppm, 10.00, 200.0ppt
   Resolution 0.01, 0.1, 1
   Accuracy ±1% F.S.
   TDS Factor 0.1 to 1.0 (default 0.5)
   Salinity  
   Range 0.00 to 80.00ppt, 0.00 to 42.00psu
   Resolution 0.01
   Accuracy ±1% F.S.
   Resistivity  
   Range 0.00 to 20.00MΩ
   Resolution 0.01
   Accuracy ±1% F.S.
   Dissolved Oxygen  
   DO Range 0.00 to 20.00mg/L
   Resolution 0.01, 0.1mg/L, selectable
   Accuracy ±0.2mg/L
   % Saturation of Oxygen 0.0 to 200.0%
   Resolution 0.1, 1%, selectable
   Accuracy ±2.0%
   Calibration Points 1 or 2 points
   Pressure Correction 60.0 to 112.5kPa/450 to 850mmHg, manual
   Salinity Correction 0.0 to 50.0g/L, manual
   Temperature  
   Range 0 to 105°C/32 to 221°F
   Resolution 0.1°C/0.1°F
   Accuracy ±0.5°C/±0.9°F
   Offset Calibration 1 point
   Calibration Range Reading ±10°C
   General Specifications  
   Temperature Compensation 0 to 100°C/32 to 212°F, manual or automatic
   Stability Criteria Low or high
   Calibration Due Alarm 1 to 31 days or off
   Auto-Power Off 10, 20 or 30 minutes after last key pressed
   Memory Stores up to 500 data sets
   Output USB communication interface
   Connector BNC, 6-pin nimi-DIN, 3.5mm jack socket
   Display Custom LCD (80×60mm)
   Power 3×1.5V AA batteries or DC5V power adapter
   Battery Life Approximately 150 hours (turn off the backlight)
   Dimensions 170(L)×85(W)×30(H)mm
   Weight 300g

  • Previous:
  • Next:

  • Write your message here and send it to us