products

Thermogravimetric analysis TGA-Q1550C

short description:


Product Detail

Product Tags

Consistent but not limited to the following national standards:

l GB∕T 14837-2018 Rubber and rubber products;

l JB/T 6856-1993 Thermogravimetric – differential thermal analyzer;

l GB/T 36246-2018 2018 Primary and middle school synthetic material surface course;

l YB/T 4328 Test method for free calcium oxide in steel slag;

l GB/T 2951.41-2008 General test method for cable and optical cable insulation and sheath materials;

Features:

l Industrial level widescreen touch structure is rich in information, including setting temperature, sample temperature, oxygen flow, nitrogen flow, differential thermal signal, various switch states, etc.

l Use the gigabit network line communication interface, the universality is strong, the communication is reliable without interruption, supports the self-recovery connection function.

l Use imported German – Sartorius balance, sensitivity of 1/100,000g.

l The furnace body is compact, temperature rise and fall speed adjustable.

l Water bath and heat insulation system, insulation high temperature furnace body temperature on the weight of the balance.

l Equipped with furnace seal system, the atmosphere can be fully protected to avoid air impact.

l Improved installation process, all adopt mechanical fixation; The sample support rod can be replaced flexibly and crucible can be matched with various models according to the requirements, so that users can have different requirements.

l The flow meter automatically switches two gas flows, fast switching speed and short stable time.

l Standard samples and charts are provided to facilitate customer calibration of constant temperature coefficient.

l Software support each resolution screen, automatically adjust the computer screen size curve display mode. Support laptop, desktop; Support Win2000, XP, VISTA, WIN7, WIN8, WIN10 and other operating systems.

l Support user edit device operation mode according to actual needs to achieve full automation of measurement steps. The software provides dozens of instructions, and users can flexibly combine and save each instruction according to their own measurement steps. Complex operations are reduced to one-click operations.

l One-piece fixed furnace body structure, without lifting up and down, convenient and safe, the rate of rising and falling can be adjusted arbitrarily.

l The removable sample holder can meet different requirements after replacement to facilitate cleaning and maintenance after sample contamination.

l The equipment adopts the cup-type balance weighing system according to the principle of electromagnetic balance.

Parameters:

1. Temperature range: RT~1550℃

2. Temperature resolution: 0.01℃

3. Temperature fluctuations: ± 0.1℃

4. Heating rate: 0.1~80℃/min

5. Cooling rate: 0.1℃/min-30℃/min(When more than 100℃, can lower the temperature at a cooling rate)

6. Constant temperature: RT-1550℃

7. Duration of constant temperature: 0-300min (When more than 1000℃, the duration is recommended to be less than 30min)

8. Temperature control mode: PID temperature control

9. Balance measurement range: 2g (optional range 10g, 50g, etc.)

10. Weight resolution: 0.01mg (optional other resolution 1ug, 0.1ug, 0.01ug, etc.)

11. *Use imported German – Sartorius balance.

12. Gas control:Nitrogen, Oxygen (automatic switching),The sealed furnace can realize nitrogen protection completely.

13. Sealed furnace body, single vent, gas generated by volatile decomposition can be collected for other equipment detection.

14. Power: 1000W, AC220V 50Hz or customize other standard power sources

15. Communication methods: Gigabit gateway communications

16. Standard crucible size(High * diameter) : 10mm*φ6mm。

17. Replaceable support, convenient for disassembly and cleaning, and can be replaced with crucible of different specifications

18. Machine size: 90cm*55cm*52cm, 90kg (110kg, with the outer packing).


  • Previous:
  • Next:

  • Write your message here and send it to us