Biological Air sampler FKC-1
Product Features
The collection port is 397 mesh with a diameter of 0.77 micropores, reducing bacterial overlap and microbial count errors
Programmable, sample size can be set arbitrarily from 1-6000L
Large sampling flow, reducing sampling time
2.2 inch TFT screen displays parameters such as sampling amount and sampling time.
Sample quantity, sampling time, and other parameters are stored on a page by page basis.
Can store 1000 sets of data
Independently developed collection chips, making sampling flow more stable
It is convenient to replace the culture dish, and the collection port can be removed to replace the culture dish (Φ90*15)
The sampling head is made of anodized aluminum oxide, suitable for most disinfection methods
Executive standards: ISO14698-1/2, GB/T 16293-2010, GMP2010.
Model Parameter FKC-1
Sampling flow 100L/min ±5%
Sampling cycle adjustment range 1-6000L (adjustable)
Control method button screen
Impact rate 10.8m/s (Anderson Impact Level 5 principle)
Battery available time over 8 hours
Overall dimension 130130160mm
G.W 7KG
Executive standards ISO 14698-1/2, GB/T 16293-2010, GMP2010